| Model Number |
3449 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/31/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device was evaluated.Visual inspection found the bic stress relief is loose.The device established communication with interfacing test fixture; however, the module does not communicate properly with the connected sensor, failure results in loss of communication to sensor.Device was observed through x-ray imaging and no defects were observed on device.The exact component failure could not be determined due to device being fully potted.
|
| |
|
Event or Problem Description
|
|
The customer reported the device worked intermittently.No patient impact or consequences were reported.
|
| |
|
Search Alerts/Recalls
|