• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION SUPERION INTERSPINOUS SPACER; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION SUPERION INTERSPINOUS SPACER; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE Back to Search Results
Model Number 100-9814
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced migration of the implant and return of the preexisting pain.Radiography confirmed a slight migration of the device to be a/p, anterior-posterior displacement.During the revision procedure, the physician attempted multiple times to loosen the implant using the driver, on the final attempt it was noted the prongs on the driver were bent.A new driver was used to successfully loosen the implant; however, the physician was unable to remove the implant.The implant was retightened, and the case aborted.The patient was referred to a spine surgeon to explant the migrated implant device.
 
Additional Manufacturer Narrative
Date of event: approximated based on the date the manufacturer became aware of the event.Additional suspect medical device component involved in the event: product family: superion instruments, upn: 102-9800, model: 102-9800, serial: n/a, batch: 204052.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERION INTERSPINOUS SPACER
Common Device Name
PROSTHESIS, SPINOUS PROCESS SPACER/PLATE
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
VERTIFLEX INC
2714 loker ave. west
carlsbad CA 92010
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key13390050
Report Number3006630150-2022-00248
Device Sequence Number12011803
Product Code NQO
UDI-Device Identifier00884662000284
UDI-Public00884662000284
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P140004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/01/2023
Device Model Number100-9814
Device Catalogue Number100-9814
Device Lot Number800040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/04/2022
Initial Report FDA Received Date01/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/02/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient RaceWhite
-
-