DEPUY SPINE INC VIPER 2 SYSTEM REDUCTION DRIVER SHAFT 5.5 X25; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 286735725
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional narrative: initial reporter is synthes sales representative without a lot number the device history records review could not be completed.Product was not returned.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the innie could not be connected to the inserter.No surgical impact and no patient impact reported.This report is for one (1) viper 2 system reduction driver shaft 5.5 x25 this is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: d10: date of concomitant therapy is december 30, 2021.H3, h6: a review of the receiving inspection (ri) for reduction shaft was conducted identifying that lot number gm3732501 was released in two batches.Batch1: released on october 25, 2012 with no discrepancies.Batch2: released on october 25, 2012 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was returned to us customer quality (cq) for evaluation.The us cq team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that spring component at the tip of the reduction shaft was found to be deformed.The observed condition was consistent as an end of life indicator for the device.No other issues where identified.After a visual inspection per guidance, it is determined that the reusable instrument device is worn from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed deformed condition of the spring component at the tip of the reduction shaft would have contributed to the complained device interaction issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument is worn from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that there was no surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER 2 SYSTEM REDUCTION DRIVER SHAFT 5.5 X25
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key13392093
Report Number1526439-2022-00159
Device Sequence Number4525090
Product Code LXH
UDI-Device Identifier10705034303499
UDI-Public(01)10705034303499
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number286735725
Device Catalogue Number286735725
Device Lot NumberGM3732501
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/03/2022
Supplement Date Received by Manufacturer02/16/2022
03/21/2022
Initial Report FDA Received Date01/31/2022
Supplement Report FDA Received Date03/17/2022
03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN LOCKING/SET SCREWS.
Patient SexUnknown
-
-