MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAB
Device Problem Priming Problem (4040)
Patient Problem Insufficient Information (4580)
Event Date 01/26/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Information received by medtronic indicated that the insulin pump was having issue during prime process.Customer stated that they were unable to exit the preparing to prime loop.No harm requiring medical intervention was reported.The device will be returned for analysis.
 
Manufacturer Narrative
Unable to perform displacement test and selftest due to motor error alarm during rewind.Unit received motor error alarm during rewind due to corroded motor home switch.Pump history download using thds was successful.No found data available date & time of alarm on pump history file on the event date.However, confirm date and time of alarm: on (b)(6) 2022 18:08:27 alarm#: 33 - alrm_broken_force_sensor.Unit was cut/open and inspected no moisture damage or component damage found inside the pump.Test reservoir lock properly inside the reservoir tube.The following were noted during visual inspection: cracked reservoir tube lip and missing endcap sticker.Confirmed motor error alarm due to corroded motor home switch during rewind.Confirmed a33 alarm at the pump history file.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAB
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key13392139
MDR Text Key295219158
Report Number2032227-2022-107305
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000643169503670
UDI-Public(01)000000643169503670
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-751NAB
Device Catalogue NumberMMT-751NAB
Device Lot NumberA4751NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received04/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/08/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age22 YR