Model Number 6054-1115S
Medical Device Problem Codes Degraded (1153); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/05/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's left hip was revised due to trunnions, gross trunnion failure, and dissociation of the head from the stem.The stem, head, and competitor liner were revised to competitor implants.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned to the manufacturer.
 
Event or Problem Description
It was reported that the patient's left hip was revised due to trunnionosis, gross trunnion failure, and dissociation of the head from the stem.The stem, head, and competitor liner were revised to competitor implants.
 
Additional Manufacturer Narrative
Reported event an event regarding disassociation involving a securfit fit stem was reported.The event was confirmed via clinician review.Method & results -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of the provided medical information by a clinical consultant indicated: what can be gleaned from the limited information provided is that a primary secure fit max 11/15 and c taper cocr head 32+5 was revised for extensive trunnionosis and dissociation of the femoral head from the stem.A single ap x-ray of the hip was provided.This demonstrated severe remodeling of the trunnion with dissociation of the femoral head.No supporting information was available to establish the root cause for this issue.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient's hip was revised due to trunnionosis, gross trunnion failure, and dissociation of the head from the stem.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY SECUR-FIT PLUS MAX #11/15
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13392145
Report Number0002249697-2022-00140
Device Sequence Number4252695
Product Code LZO
UDI-Device Identifier07613327020908
UDI-Public07613327020908
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K041170
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2010
Device Model Number6054-1115S
Device Catalogue Number6054-1115S
Device Lot Number30579501
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date01/31/2022
Supplement Report FDA Received Date05/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/06/2005
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age67 YR
Patient SexUnknown
-
-