| Model Number |
RLT231414 |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
01/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An imaging evaluation was conducted.The evaluation found: there are 2 limbs within the r common iliac.The ipsi limb extension provided by the second device is <1cm.There appears to be >3cm of device implanted within the r common iliac.The diameter near the aortic bifurcation 15.3mm.This would be better measured on a pre-implant scan but these images were not provided.No additional scans demonstrating procedural details were provided that would indicate the prevention of the observed occlusion related to a possibly <18mm distal waist or placing 2 devices vs 1 in the distal waist.(possible procedural details not provided: e.G.Simultaneous ballooning at the distal waist or the use of a balloon expandable stent to reinforce the single ipsilateral limb) outside of ifu.There is 1 contralateral limb implanted which is compressed at the area of the distal waist.The flow divider is not compressed.No contrast is visualized within the contralateral limb.Contrast does reconstitute distally providing perfusion to the scanned portion of the l external and common femoral artery.This event also includes device (b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021, the patient was treated for an abdominal aortic aneurysm.The physician implanted a gore® excluder® aaa endoprosthesis featuring c3® delivery system.He also implanted a gore® excluder® aaa endoprosthesis on the left side, but it was about 1 cm short of the lia, so he implanted a second gore® excluder® aaa endoprosthesis to reach to the lia.The patient tolerated the procedure.On (b)(6) 2022, the patient was treated for an obstruction on the left side of the main body.It was found that the left limb had collapsed from the flow divider on down.The physician believes this was due to the two plc devices that had been implanted.The limbs were crossed, so that in addition to the fact that the vessel was heavily calcified and of small diameter, it resulted in pressure that caused the limb to collapse.A thrombectomy was performed, and two gore® excluder® aaa endoprostheses were implanted to reopen the limb.The patient tolerated the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Added g3/g4, h1/h2, the stent graft trunk ipsilateral leg c3 (rlt231414) did not cause any injury and the patient tolerated the procedure.It is a non-complaint.The stent graft contralateral leg (plc141000 and plc141200) caused the compression in the patient.
|
| |
|
Search Alerts/Recalls
|