C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER Back to Search Results
Model Number 998706
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the catheter balloon burst during inflation.
 
Additional Manufacturer Narrative
The reported event is confirmed, cause unknown.The x-force balloon dilation catheter was returned without the original packaging.An evaluation was completed by futurematrix on 19apr2022.The sample was visually inspected at 12-18" under normal room lighting; the outer shaft showed no defects and none were noted with the hubs assembly.Portholes, bullets and markers were all present and no non-conformities were noted.An in-house 0.038" guidewire was used and passed freely per specifications.Further examination of the balloon was conducted under magnification; a hole was noted mid-barrel of the balloon.The balloon was dissected away from the shaft of the catheter; the balloon was removed from the device and cut open to be examined from the inside out.A longitudinal tear was noted from the inside of the base balloon which caused the fiber separation.All inspection data was found to be within specification.It is unknown if a prep inflation occurred prior to the procedure and/or if the inflation device was used or if exceeded rpb; therefore, the actual root cause is unknown.A potential root cause for this event could be, "poor balloon design (molecular orientation, wall thickness), inadequate material selection, improper fiber orientation".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "do not exceed the recommended rated burst pressure (rbp) for this device.Balloon rupture may occur if the rbp rating is exceeded.Please refer to the device label for the recommended maximum rated burst pressure.Extreme caution should be used when considering dilation of irregular, not-compliant tissue or in the presence of certain calculi." "only a physician who has an understanding of the clinical applications, technical principles and associated risks of balloon dilation within the urinary tract should use this device." "all x-force® balloon dilation catheters contain air in the balloon lumen.The air must be removed to allow liquid to fill the balloon when it is inflated." "inflating the balloon catheter 1.Fill the inflation device (eagle¿ inflation device) with sterile diluted contrast medium.2.Attach the inflation device to the balloon lumen.3.Inflate the balloon "note: do not use air or any gaseous substances as a balloon inflation medium always use sterile liquid media." "note: the use of an inflation device with a pressure gauge is highly recommended to make sure adequate pressure is applied and the rated burst pressure of the balloon is not exceeded." "caution: if loss of pressure in the balloon occurs or the balloon ruptures, immediately stop the procedure, deflate the balloon and remove carefully.Do not attempt to re-inflate the balloon.If after inspection it is noted pieces of the balloon are missing, check for and endoscopically retrieve the fragments when possible." the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the catheter balloon burst during inflation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® BALLOON DILATION CATHETER
Common Device Name
BALLOON DILATION CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13392477
Report Number1018233-2022-00246
Device Sequence Number19896810
Product Code EZN
UDI-Device Identifier10801741052740
UDI-Public(01)10801741052740
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K050875
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/30/2023
Device Model Number998706
Device Catalogue Number998706
Device Lot NumberBMFSFM30
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/10/2022
Supplement Date Received by Manufacturer05/20/2022
Initial Report FDA Received Date01/31/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-