MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTAPRO; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTAPRO; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 3243
Medical Device Problem Codes Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that burr separation occurred.A 1.75mm rotapro was selected for the right coronary artery (rca) procedure.While the physician adjusting the rotation speed to 180,000rpm, the rotapro burr detached outside the patient.The procedure was completed with another rotapro device.There were no complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the returned product consisted of the rotapro atherectomy device.The device was received with the burr separated.The burr was not returned for analysis.The advancer, drive shaft, handshake connection, and coil were visually and microscopically examined.Inspection of the device found that the coil had been kinked and stretched at the point of detachment from the burr.Product analysis confirmed the reported event, as the coil was kinked, stretched and broken.
 
Event or Problem Description
It was reported that burr separation occurred.A 1.75mm rotapro was selected for the right coronary artery (rca) procedure.While the physician adjusting the rotation speed to 180,000rpm, the rotapro burr detached outside the patient.The procedure was completed with another rotapro device.There were no complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road cork
cork IRELA ND
EI   IRELAND
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13393367
Report Number2134265-2022-00898
Device Sequence Number14516313
Product Code MCX
UDI-Device Identifier08714729893370
UDI-Public08714729893370
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
P900056
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/27/2023
Device Model Number3243
Device Catalogue Number3243
Device Lot Number0027919823
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/26/2022
Supplement Date Received by Manufacturer03/15/2022
Initial Report FDA Received Date01/31/2022
Supplement Report FDA Received Date03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-