ARTHREX, INC. COMPR FT SCRW,3.5 MINI, 42MM; PLATE, FIXATION, BONE Back to Search Results
Model Number COMPR FT SCRW,3.5 MINI, 42MM
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a metal removal of two screws ar-8730-42hs (lot 10327710) and ar-8730-46hs (lot 10327712) the head of the screw ar-8730-46hs (lot 10327712) broke during unscrewing with the normal screwdriver for the device.The device ar-8737-62 (lot 1391931) was then used to grasp the broken screw and also broke.The broken screw had to be chiseled free to remove it, which caused much damage to the bone.The duration of the surgery was prolonged by approximately 45 minutes due to the failures.No broken parts remained inside the patient.The surgery was finished successfully and it was not necessary to do a second surgery.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Complaint confirmed, the drive feature was found to be deformed and cracked.The threads were also damaged and the tip fragment of the ar-8737-62 was observed in the cannulation at the proximal end.The cause of the event is undetermined, however likely causes included user-applied mechanical forced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPR FT SCRW,3.5 MINI, 42MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13396097
Report Number1220246-2022-04355
Device Sequence Number5893143
Product Code HRS
UDI-Device Identifier00888867303997
UDI-Public00888867303997
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K201132
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberCOMPR FT SCRW,3.5 MINI, 42MM
Device Catalogue NumberAR-8730-42HS
Device Lot Number10327710
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer01/12/2022
Initial Report FDA Received Date01/31/2022
Supplement Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/25/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-