• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. UNIVERSAL GLENOID INLAY SMALL; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. UNIVERSAL GLENOID INLAY SMALL; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number UNIVERSAL GLENOID INLAY SMALL
Health Effect - Clinical Code Metal Related Pathology (4530)
Date of Event 01/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that a revision surgery of a eclipse with universal glenoid was performed.No further information was provided.Update 19-jan-2022: further information were provided that the revision surgery was necessary due to an aseptic loosening.It was further reported that a progressive polyethylene abrasion with pronounced metallosis was identified as well.Therefore the shoulder prosthesis was changed to an inverse shoulder prosthesis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL GLENOID INLAY SMALL
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13397023
Report Number1220246-2022-04359
Device Sequence Number5893144
Product Code PHX
UDI-Device Identifier00888867057418
UDI-Public00888867057418
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K182039
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberUNIVERSAL GLENOID INLAY SMALL
Device Catalogue NumberAR-9121-01
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date01/13/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2022
Initial Report FDA Received Date01/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-