| Model Number |
DRIVER SHAFT, T25 HEXALOBE, CANNULATED |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a screw removal the driver shaft became twisted and snapped off during the removal.There was no harm for patient, operator or third party reported.The surgery was finished successfully.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint is not confirmed.One unpackaged ar-8610d-65 was received for investigation.Visual inspection under the aid of micro-vu id: 654 concluded without the observance of any breakage across the drive geometry (investigation photos #5-6).No problem found.
|
| |
|