SYNTHES GMBH DRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA; REAMER Back to Search Results
Catalog Number 314.743
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional narrative: additional device product codes: hrx.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6), 2022, the drill bit of the estuary hose split after the graft extraction finished.This report involves one (1) drive shaft-minimum 520mm length-for use with ria.This is report 1 of 1 for (b)(4).
 
Event or Problem Description
It was further reported that no broken fragments retained in the patient.The procedure was completed successfully with no surgical delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6: investigation summary: the complaint device (ria driveshaft l520) was not received for investigation.A photo investigation was performed based on the photo provided.The image was reviewed, and the complaint condition is confirmed.The tip of the device is broken, fragment is missing in the evidence provided.A definite assignable root cause could not be determined based on the provided information.As the device was not returned, and as-received condition could not be assessed and a dimensional inspection and document/specification review were not completed.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.H3, h6: a device history record (dhr) review was conducted: part # 314.743.Synthes lot # h427158.Supplier lot # h427158.Release to warehouse date: 23 oct 2018.Supplier: (b)(4).No ncr's were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that ria driveshaft l520, 314.743/h427158 was found broken at the distal end, the broken piece was not returned.Additionally, the ria driveshaft l520 is bent from the shaft.The dimensional inspection at the point of breakage was performed, and it is within the specification limit.The observed condition of the ria driveshaft l520 was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the ria driveshaft l520 was observed to be broken and bent.While no definitive root cause could be determined, it is probable that the ria driveshaft l520 was broken due to unintended forces.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA
Common Device Name
REAMER
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13397879
Report Number8030965-2022-00605
Device Sequence Number17274390
Product Code HTO
UDI-Device Identifier07611819739208
UDI-Public(01)07611819739208
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K042899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number314.743
Device Lot NumberH427158
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/03/2022
Supplement Date Received by Manufacturer02/04/2022
02/11/2022
Initial Report FDA Received Date01/31/2022
Supplement Report FDA Received Date02/07/2022
03/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/23/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1