WUXI JIAJIAN MEDICAL INSTRUMENT CO., LTD METRON SELF ADHESIVE ELECTRODES; ELECTRODE, CUTANEOUS Back to Search Results
Model Number 32028582
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 10/27/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Nmt purchases the device from (b)(4).The receipt of the original complaint was on (b)(6) 2021 stated "there was no serious injury or death.This was a minor injury that needed minor treatment.They put a bandage over the wound.The patient did not see a dr only a physical therapist.This physical therapist was a traveling therapist who has already left the facility.The customer noted she can reach out if more information in needed from that physical trainer.The type of treatment that was being used was interferential.(b)(6) stated the ifu for product is assumed to have been used.The intensity was at 61 with the length of time of 15 minutes.Therapist can treat under scope of practice not by order.The patient did discontinued use of this product.The equipment was taken out of service as well.Patient mentioned injury a day later, was not initially apparent.The wound was in the general area of where the equipment was used." this information was forwarded to (b)(4) on (b)(6) 2021 for investigation.(b)(4) requested additional information, which was provided by nt including the unit the electrode was used with: "the unit was a dynatronic ultrasound stimulation unit (model#d708 / serial#(b)(4)).Manufacture date was 10/2003.Let me know if there is any other information that is needed." nmt was provided the fda medwatch report on 1/21/2022 with more details of the " burn" and this report was provided to (b)(4) on (b)(6) 2022.This is when nmt determined that as the importer, the event was reportable.To date (b)(4) has offered the following, "we think it could be low temperature burn, it happened because there is no heat dissipation during the treatment.This is only our conjecture." no samples have been returned for evaluation or photos.
 
Event or Problem Description
On (b)(6) 2021, the first report of an issue with tns metron electrode 2" square was reported, however very little information was given.On (b)(6) 2021, it was reported to (b)(6) customer service by (b)(6) that "there was no serious injury or death.This was a minor injury that needed minor treatment.They put a bandage over the wound.The patient did not see a dr only a physical therapist.This physical therapist was a traveling therapist who has already left the facility.The customer noted she can reach out if more information in needed from that physical trainer.The type of treatment that was being used was interferential.(b)(6) stated the ifu for product is assumed to have been used.The intensity was at 61 with the length of time of 15 minutes.Therapist can treat under scope of practice not by order.The patient did discontinued use of this product.The equipment was taken out of service as well.Patient mentioned injury a day later, was not initially apparent.The wound was in the general area of where the equipment was used." with this original information it was determined that the event was not reportable at this time due no serious injury or death.However on (b)(6) 2022, nmt was made aware of that an fda medwatch report was initiated by the customer which stated the patient had a " single, small, circular burn-like sore" from use of the device.With this information, nmt deemed the event to be reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRON SELF ADHESIVE ELECTRODES
Common Device Name
ELECTRODE, CUTANEOUS
Manufacturer (Section D)
WUXI JIAJIAN MEDICAL INSTRUMENT CO., LTD
no. 35 baiqiao rd., ehu town
xishan district,, wuxi,
jiangsu, china 21411 6
CH  214116
MDR Report Key13399329
Report Number1317188-2022-00002
Device Sequence Number20306581
Product Code GXY
UDI-Device Identifier10885914070755
UDI-Public10885914070755
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/31/2022
Report Date (Section F) 11/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model Number32028582
Device Catalogue NumberTNS METRON ELECTRODE 2" SQUARE
Device Lot Number172026
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/21/2022
Device Age3 YR
Event Location Outpatient Treatment Facility
Date Report to Manufacturer01/24/2022
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date01/31/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
Date Report Sent to FDA11/01/2021
-
-