(b)(4).Date of event: unknown; captured as awareness date.Batch # 348a55.Device analysis: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the tcr75 cartridge was received with no apparent damage and with no instrument.The swing tab was found to be in the unlocked position, unfired and with one missing staple along the cartridge deck.The reload was tested for functionality with a test device and it fired, cut, and formed all the staples as intended.The staple line and cut line were complete, and the staples meet the staple form release criteria.In addition a tyvek was received along with the device.It is possible that improper handling of the reload caused the staple to fell out, the proper handling instructions should be used when removing and reloading cartridges into the device, please reference the instructions for use.It should be noted that 100% inspection is performed by means of an automated vision system to ensure staples presence; the swing tab is present and unlocked.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|