| Model Number |
7207568 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference case (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy procedure, a biorci screw was used to fix the graft in the tibial tunnel and when screwing it broke; all pieces of the broken screw were removed.The procedure was successfully completed with non-significant delay using a back-up device.No further complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy procedure, a biorci screw was used to fix the graft in the tibial tunnel and when screwing it broke; all pieces of the broken screw were removed with a screwdriver and nitinol guide leaving no residue inside the patient.The procedure was successfully completed with non-significant delay using a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was received for evaluation.It was determined the device did not contribute to the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the resin found that the storage protocols, material specifications, and material tests were appropriately documented.A material certificate of analysis is required.A review of the customer provided image found an image of a broken screw.There is debris on the device.A visual inspection of the returned device found that it is not in its original packaging.The screw is fractured in the center and on the distal end.There is debris in the threads.Based on the condition of the product material found during visual inspection, additional material testing is not required.Per case details, the broken pieces were retrieved from the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
| |
|