As reported, during an atherectomy procedure involving the right superficial femoral artery, a flexor high-flex ansel guiding sheath unraveled and separated upon removal of the device.The device was flushed and inserted over a bentson wire through the left common femoral artery.Resistance was encountered upon initial advancement.The anatomy was reportedly heavily calcified around the junction of the internal and external iliac arteries.Upon meeting resistance, the user stopped advancing the sheath and injected contrast to visualize under fluoroscopy.During attempted removal of the device, the shaft unraveled and separated, leaving approximately five to eight centimeters of the sheath in the patient.The dilator and wire guide were in the lumen of the sheath at the time of separation.The user attempted to snare the separated fragment with multiple devices but was unsuccessful.The patient was transferred from the outpatient center to the hospital across the street for surgical removal.The separated fragment was successfully removed during an endarterectomy with patch repair.
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Event summary as reported, during an atherectomy procedure involving the right superficial femoral artery, a flexor high-flex ansel guiding sheath unraveled and separated upon removal of the device.The device was flushed and inserted over a bentson wire through the left common femoral artery.Resistance was encountered upon initial advancement.The anatomy was reportedly heavily calcified around the junction of the internal and external iliac arteries.Upon meeting resistance, the user stopped advancing the sheath and injected contrast to visualize under fluoroscopy.During attempted removal of the device, the shaft unraveled and separated, leaving approximately five to eight centimeters of the sheath in the patient.The dilator and wire guide were in the lumen of the sheath at the time of separation.The user attempted to snare the separated fragment with multiple devices but was unsuccessful.The patient was transferred from the outpatient center to the hospital across the street for surgical removal.The separated fragment was successfully removed during an endarterectomy with patch repair.Investigation - evaluation reviews of the complaint history, device history record, instructions for use, manufacturing instructions, and quality control procedures and a visual inspection of the device were conducted during the investigation.One flexor high-flex ansel guiding sheath was returned for investigation in a used and damaged condition.The distal section of the sheath was separated 47 cm from the white cap on the proximal end.The tip section was not returned.A document-based investigation evaluation was performed.No related non-conformances were recorded, and there have been no other reported complaints for this lot number.The device history record review provides objective evidence that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.A review of relevant manufacturing documents was conducted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The device is provided with instructions for use which warn, ¿if resistance is encountered during advancement of flexor sheath, assess cause of resistance and consider dilation of any restriction identified or consider alternate treatment strategy.If flexor sheath is advanced through resistance, force to remove the sheath will be higher, increasing the risk of sheath material or hub separation upon withdrawal.Reinsertion of dilator prior to removal of flexor sheath increases the strength of the sheath and lessens the risk of device separation.If resistance is anticipated or encountered during withdrawal of flexor sheath, consider reinserting the dilator prior to continuing removal.¿ based on the available information, cook has concluded that the patient¿s heavily calcified anatomy contributed to the reported resistance when advancing the sheath and to the sheath separation.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|