• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDLINE INDUSTRIES LP; TTL1LYR 16FR10ML 100%SIL TMPST TR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDLINE INDUSTRIES LP; TTL1LYR 16FR10ML 100%SIL TMPST TR Back to Search Results
Catalog Number URO180816TS
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that the catheter was leaking so the nurse checked the balloon and there was only 8ml in the balloon.The nurse re-inflated the balloon to ensure that the balloon had a total of 10ml in the balloon however the catheter continued to leak so the nurse removed the foley catheter and inserted a new foley catheter.The nurse noted that the second catheter was fully inflated and seated in the bladder however the leaking continued so the catheter was removed.It was not reported that an additional catheter was placed.The catheter was returned to the manufacturer for evaluation and the customer reported issue of a leaking foley catheter balloon could not be confirmed through functional evaluation of the received sample.A definitive root cause for the reported issue could not be determined at this time.There is no additional information available.Due to the reported incident and in an abundance of caution, this medwatch is being filed.If additional relevant information becomes available a supplemental medwatch will be filed.
 
Event or Problem Description
The customer reported that the foley catheter balloon was leaking so the nurse removed the foley catheter and replaced the device with a new catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
TTL1LYR 16FR10ML 100%SIL TMPST TR
Manufacturer (Section D)
MEDLINE INDUSTRIES LP
three lakes drive
northfield IL 60093
Manufacturer Contact
karen trutsch
three lakes drive
northfield, IL 60093
MDR Report Key13404904
Report Number1417592-2022-00007
Device Sequence Number17665027
Product Code OHR
UDI-Device Identifier10889942127955
UDI-Public10889942127955
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberURO180816TS
Device Lot Number21CBE450
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/04/2022
Initial Report FDA Received Date01/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-