• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. HELICAL-10-AV; ARTIFICIAL EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. HELICAL-10-AV; ARTIFICIAL EMBOLIZATION DEVICE Back to Search Results
Model Number 100515HCSR-R-V-A1
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during aneurysm embolization procedure, when the coil was deployed in the aneurysm resistance encountered and only half the coil deployed.The coil unintentionally detached in the microcatheter when it was being removed from the patient.The coil along with the microcatheter were successfully removed and procedure completed successfully.There was no report of harm or injury to the patient.
 
Additional Manufacturer Narrative
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was returned to the manufacturer for analysis.The investigation is currently underway.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with use of the device.
 
Additional Manufacturer Narrative
Additional information: h6, h10 (summary device evaluation).Summary device evaluation: the microcatheter used in the procedure was not evaluated as a part of this evaluation, so the investigation could not determine if it had caused or contributed to the reported complaint.The implant was the only component returned for evaluation.The investigation of the returned implant found the loops to be deformed at proximal.The stretch resistance monofilament was cut, and the implant regained shape.The shape loss appears to be the result of an increased tension in the monofilament - this issue is tracked by quality and manufacturing.The coil pusher was not returned for evaluation.Without the returned and evaluation of the pusher, the investigation is unable to determine the mode of separation (i.E.Unintentional vs normal detachment using a detachment controller); however, the implant was found with a tensile break mono shape at the proximal marker band which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HELICAL-10-AV
Common Device Name
ARTIFICIAL EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise drive
aliso viejo, CA 92656
7142478000
MDR Report Key13405191
Report Number2032493-2022-00036
Device Sequence Number4935957
Product Code HCG
UDI-Device Identifier00812636027531
UDI-Public(01)00812636027531(11)170627(17)220531(10)17062753A
Combination Product (Y/N)N
Initial Reporter CountryKS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model Number100515HCSR-R-V-A1
Device Catalogue Number100515HCSR-R-V
Device Lot Number17062753A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer01/28/2022
Initial Report FDA Received Date01/31/2022
Supplement Report FDA Received Date02/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age50 YR
Patient SexFemale
Patient Weight62 KG
-
-