Model Number 9-VSD-MUSC-016
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event or Problem Description
It was reported that a 16mm amplatzer muscular vsd occluder was selected for an implant on (b)(6) 2022.The device was navigated through non abbott delivery sheath during deployment.However the physician didn't achieved the desired shape.It shape like a cobra head.Device was removed from the patient prior to release from the delivery cable.There was no interaction with atrial structures during deployment and no angulation or kink noticed in the delivery system.No device was implanted, patient stable, no additional information was provided.
 
Additional Manufacturer Narrative
The reported event of deformation of a 16mm amplatzer muscular vsd occluder could not be confirmed.The investigation confirmed the device met functional specifications when analyzed at abbott under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER MUSCULAR VSD OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key13406277
Report Number2135147-2022-00049
Device Sequence Number1843072
Product Code MLV
UDI-Device Identifier00811806011844
UDI-Public00811806011844
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P040040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model Number9-VSD-MUSC-016
Device Catalogue Number9-VSD-MUSC-016
Device Lot Number5862564
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2022
Supplement Date Received by Manufacturer03/10/2022
Initial Report FDA Received Date01/31/2022
Supplement Report FDA Received Date03/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/16/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-