The user facility reported that during a below-the-knee angioplasty procedure the circulating technician opened the ga1430 and the scrub tech flushed the holder.Upon passing the device to the physician, it was noted that the tip of the wire was damaged.The wire was set aside and a new ga1430 was opened and prepped, and procedure continued without issue.The procedure was completed successfully, and the patient was transferred to the recovery area.The estimated blood loss was less than 250cc's.There were no other devices or equipment used with the reported product.
|
|
This report is being submitted as follow up no.1 to provide the device return date in section d9, update sections h3 and h4, and to provide the completed investigation results.The actual sample was received by ashitaka factory on january 31, 2022.Visual inspection of the actual sample found that the core wire was exposed at the distal end.Magnifying inspection of the actual sample obtained the following results.The outer layer and the gold coil were missing from the distal end.The outer layer had been constricted near the fracture end.The fracture surface of the outer layer was in an oval.Gold coil was confirmed inside the outer layer.Electron microscopic inspection of the actual sample obtained the following results.The fracture end seemed to have been torn-off, and some creases were observed on the outer surface of the distal end section.On the distal end of the core wire, there was an evidence of sizing cut performed in our manufacturing process.From this, it was most unlikely that there was a portion missing from the core wire.The outer diameter of the actual sample was measured and confirmed to meet the control standards.From the condition of the actual sample, it was presumed that some tensile load was applied to the distal end.Assuming the possibility that the actual sample was taken out of the holder tube with resistance (e.G., with a clip hub attached to it), a simulation test was conducted.As a result, the following characteristics were observed on the test sample.The outer layer and the gold coil were missing from the distal end.Creases were observed on the fractured outer layer surface, and the fracture end had been torn off.These characteristics were conceivable to be similar to those observed on the actual sample.Based on the results of the investigation, as a possible cause of occurrence, it is likely that a tensile load was applied to the distal end of the actual sample due to some factor when it was pulled out from the holder tube, which resulted in the missing outer layer and gold coil.As for the core wire, it is considered that there was no portion missing from the core wire since the evidence of sizing cut performed in our manufacturing process was confirmed on the distal end.Ashitaka factory assures in our manufacturing process, 100% visual inspection is performed after the packaging process to assure no damage on the distal end.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.9681834.
|