Model Number 5829502
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code Extravasation (1842)
Date of Event 01/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.(expiry date: (b)(6) 2019).
 
Event or Problem Description
It was reported that post port placement procedure, the extravasation on the lateral side was allegedly found.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that post port placement procedure, the extravasation on the lateral side was allegedly found.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one powerport mri duo attached to a catheter was returned for evaluation and two electronic photos were provided for review.Visual, microscopic visual and functional evaluation were performed on the returned device.The investigation is confirmed for the identified material protrusion and reported fluid leak issue as one of the two port septa(right) was partially dislodged and upon infusing the port, leak was noted from the partially dislodged septum.Further during functional evaluation mandrel test was performed were both the port stem passed the mandrel test as the black mark fully inserted into the lumen and the radiographic images shows no blockage of lumens in the region of port stem and the lumens appeared clear.A definitive root cause for the reported failure could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 12/2019) h6 (device, method).H11: h6 (result, conclusion).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that approximately three years and six months post port placement procedure, the extravasation on the lateral side was allegedly found.It was further reported that, the device was removed on later days.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13410828
Report Number3006260740-2022-00122
Device Sequence Number17111744
Product Code LJT
UDI-Device Identifier00801741027406
UDI-Public(01)00801741027406
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K090512
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5829502
Device Catalogue Number5829502
Device Lot NumberRECQ0491
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer02/22/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date02/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/06/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-