(b)(4).On (b)(6) 2022 the customer passed.The pump passed the functional test, including the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and the dat test at 0.0869 inches.The following were noted during visual inspection: minor scratched display window, missing display window cover, scratched case, cracked battery tube threads, cracked case at battery tube side, pillowing keypad overlay, cracked retainer and corroded battery tube.The test p-cap and reservoir does lock in place in the reservoir compartment.History download was successful using thus and carelink upload was successful.The pump did not have a battery installed when received.Please see below for the date range listed in the formatted history file.The formatted history file lists data from 06/29/2020 to 12/23/2021.The pump passed the functional testing.During visual inspection the pump had a corroded battery tube.Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
Updated analysis summary by (b)(4) on apr 15, 2022 retainer ring = clear.On jan 01, 2022 the customer passed.No allegation to pump defectiveness or bg issue noted.Device passed the functional test, including the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and the delivery accuracy test at 0.0869 inches.The following were noted during visual inspection: minor scratched display window, missing display window cover, scratched case, cracked battery tube threads, cracked case at battery tube side, pillowing keypad overlay, cracked retainer and corroded battery tube.The test p-cap and reservoir does lock in place in the reservoir compartment.History download was successful using thus and carelink upload was successful.Device did not have a battery installed when received.Please see below for the date range listed in the formatted history file.The formatted history file lists data from 06/29/2020 to 12/23/2021.Device passed the functional testing.During visual inspection the pump had a corroded battery tube.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|