C.R. BARD, INC. (BASD) -3006260740 HICKMAN 9F DUAL LUMEN CATHETERS; CHRONIC CATHETERS Back to Search Results
Model Number 0600600
Medical Device Problem Codes Burst Container or Vessel (1074); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiration date: (b)(6) 2025).
 
Event or Problem Description
It was reported that two months and twenty one days post catheter placement, the catheter was allegedly fractured.It was further reported that catheter needs to be removed due to fracture.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that two months and twenty one days post catheter placement, the catheter was allegedly fractured.It was further reported that catheter needs to be removed due to fracture.There was no reported patient injury.
 
Event or Problem Description
It was reported that two months and twenty one days post catheter placement, the catheter was allegedly fractured.It was further reported that catheter need to be removed due to fracture.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one 9fr hickman d/l catheter was returned for evaluation and two electronic photos were provided for review.Gross, microscopic visual, functional testing and tactile evaluation were performed.The investigation is unconfirmed for the reported fracture issue as burst was identified during return sample evaluation.However, the investigation is confirmed for the identified burst issue as a s-shape split was noted on the catheter body approximately 0.4cm from the distal end of the bifurcation.Also, a complete circumferential break was noted to the distal end of the catheter and striations were observed throughout the catheter.Furthermore, the edges of the s-shaped split on the catheter appeared to have a smooth break surface.Upon infusion of the white luer, leaking was noted from the s-shaped split on the catheter.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 01/2025), g3, h6(device).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HICKMAN 9F DUAL LUMEN CATHETERS
Common Device Name
CHRONIC CATHETERS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13412235
Report Number3006260740-2022-00125
Device Sequence Number4690544
Product Code LJS
UDI-Device Identifier00801741051890
UDI-Public(01)00801741051890
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K830233
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0600600
Device Catalogue Number0600600
Device Lot NumberHUEY1756
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer05/09/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date05/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexMale
-
-