Model Number 139005
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tigertail ureteral catheters had the syringe connector screwed on so tight that it was damaging the catheters and not allowing the wire to pass through them.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the tigertail ureteral catheters had the syringe connector screwed on so tight that it was damaging the catheters and not allowing the wire to pass through them.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned for evaluation.A potential root cause for this event could be, "incorrect geometry".The device was used for treatment purposes, however, it was unknown if the device had met relevant specifications or contributed to the reported event.The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "after placement of the cystoscope into the bladder, the ureteral orifice is located and examined and a choice of ureteral catheter is made.Under direct vision, the guidewire is positioned within the ureteral orifice." "the ureteral catheter is inserted over the guidewire and advanced into the ureter.The guidewire is removed and the catheter adapter is secured to the catheter.Gentle aspiration to evacuate air from the catheter can be performed at this time.Contrast media is then injected through the ureteral catheter into the ureter.Under fluoroscopy, the ureter is visualized and required intervention determined." "caution: do not over-tighten the catheter adapter.Over-tightening the catheter adapter may occlude the lumen of the catheter." h11:section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® TIGERTAIL¿ FLEXIBLE TIP URETERAL CATHETERS
Common Device Name
URETHRAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13415452
Report Number1018233-2022-00265
Device Sequence Number3871469
Product Code KOD
UDI-Device Identifier10801741127097
UDI-Public(01)10801741127097
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K033719
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number139005
Device Catalogue Number139005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/10/2022
Supplement Date Received by Manufacturer04/28/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-