• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. JGRKNT 1.0MM MINI 2-0 NDLS; FASTNER, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. JGRKNT 1.0MM MINI 2-0 NDLS; FASTNER, FIXATION Back to Search Results
Model Number 912076
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: foreign: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that the sleeve did not move on the instrument as intended.The procedure was completed with another product.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Visual examination of the returned product identified the anchor and suture line attached to the needle tip.A function test was performed and the device functioned as intended.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Upon conclusion of the investigation, no problem was found with the given device.It was determined to be not reportable.The initial report was forwarded in error and should be voided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JGRKNT 1.0MM MINI 2-0 NDLS
Common Device Name
FASTNER, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13415709
Report Number0001825034-2022-00240
Device Sequence Number5232526
Product Code DZL
UDI-Device Identifier00880304520820
UDI-Public(01)00880304520820(17)251120(10)836410
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K110879
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/20/2025
Device Model Number912076
Device Catalogue Number912076
Device Lot Number836410
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer03/24/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date03/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-