Catalog Number 04P75-03
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On 25-jan-2022, abbott point of care (apoc) was contacted by a customer who reported observing smoke coming from the compartment of i-stat analyzer (b)(4) during operation.The product did not become hot to the touch.There was no additional information at the time of this report.There were no injuries reported.The analyzer was replaced at no charge and returning for investigation.Apoc has determined that a component failure within the analyzer circuitry, may lead to the batteries becoming uncomfortably hot to touch in the area of the battery compartment when using a green non-fused battery carrier.The analyzer was operating on rechargeable batteries at the time of the event.At this time there is no reason to believe product will become hot to the touch.The product was replaced and returning for investigation.Based on the information available, there were no patient or user related injuries associated with this complaint.
 
Additional Manufacturer Narrative
Apoc incident:(b)(4).The investigation was completed on 12-jul-2022.The customer reported that smoke was emitted from the battery compartment of analyzer s/n (b)(6) during operation.The customer also reported the analyzer was warm when the smoke was observed and they were using an i-stat rechargeable battery.A subsequent attempt to activate the analyzer using a different battery was unsuccessful.Failure analysis confirmed the complaint of inability to activate the analyzer, and the cause was determined to be the failure of component tv4 on the battery contact pcb.This is also the likely source of the smoke that was reported by the customer, as tv4 was found to be burnt.The rechargeable battery that was in use at the time of the incident was not returned, and therefore cannot be ruled out as a contributing factor.
 
Event or Problem Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT1 ANALYZER, IMMUNO READY, WIRELESS
Common Device Name
I-STAT1 ANALYZER
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
FLEXTRONICS MANUFACTURING (SINGAPORE)
pte ltd
1 kallang place
singapore 33921 1
SN   339211
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key13415716
Report Number2245578-2022-00019
Device Sequence Number20141814
Product Code CGA
UDI-Device Identifier00054749005529
UDI-Public00054749005529
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K001387
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04P75-03
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/25/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date07/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-