| Model Number |
N/A |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient had a total hip arthroplasty.The patient was non-compliant (due to the patient not following restrictions at rehab) and the cup, augment block, and screws loosened.Therefore, the patient was revised with a bigger cup, bigger augment, and screws approximately 1-2 months later.Subsequently, the patient was non-compliant again, therefore will be revised again in the near future with a custom trifflange.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant products: cat# 110010267 g7 osseoti multihole 58mm g lot#7084562; cat# 802804004 zb type 1 ceramic hd 40mm x +6 lot# 3045483; cat# 51-101070 tprlc 133 fp type1 pps ho 7.0 lot#6656562; cat# 00625006535 bone scr 6.5x35 self-tap lot# j7105903; cat# 00625006540 bone scr 6.5x40 self-tap lot# j7031156; cat# 00625006540 bone scr 6.5x40 self-tap lot# 65094425; cat# 00625006540 bone scr 6.5x40 self-tap lot# j7130462; cat# 00625006550 bone scr 6.5x50 self-tap lot# j6909648; cat# 00625006550 bone scr 6.5x50 self-tap lot# j6909648; cat# 00489405020 trabecular metalâ¿¢ acetabular revision system acetabular augment lot# 64634283.The device will not be returned for analysis, as the device remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2022 - 00261, 0001822565 - 2022 - 00262, 0002648920 - 2022 - 00021, 0001822565 - 2022 - 00263, 0001822565 - 2022 - 00264.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: d4; h2; h3; h4; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were not provided.A review of the device history records identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to the patient being non compliant to post op care as reported by the surgeon.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|