• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. BONE SCR 6.5X50 SELF-TAP; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. BONE SCR 6.5X50 SELF-TAP; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had a total hip arthroplasty.The patient was non-compliant (due to the patient not following restrictions at rehab) and the cup, augment block, and screws loosened.Therefore, the patient was revised with a bigger cup, bigger augment, and screws approximately 1-2 months later.Subsequently, the patient was non-compliant again, therefore will be revised again in the near future with a custom trifflange.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: cat# 110010267, g7 osseoti multihole 58mm g, lot#7084562.Cat# 802804004, zb type 1 ceramic hd 40mm x +6.Lot# 3045483.Cat# 51-101070, tprlc 133 fp type1 pps ho 7.0, lot#6656562.Cat# 00625006535, bone scr 6.5x35 self-tap, lot# j7105903.Cat# 00625006540, bone scr 6.5x40 self-tap, lot# j7031156.Cat# 00625006540, bone scr 6.5x40 self-tap, lot# 65094425.Cat# 00625006540, bone scr 6.5x40 self-tap, lot# j7130462.Cat# 00625006550, bone scr 6.5x50 self-tap, lot# j6909648.Cat# 00625006550, bone scr 6.5x50 self-tap, lot# j6909648.Cat# 00489405020, trabecular metala acetabular revision system acetabular augment, lot# 64634283.The device will not be returned for analysis, as the device remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-00261, 0001822565-2022-00262, 0002648920-2022-00021, 0001822565-2022-00263, 0001822565-2022-00265.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: d4; h2; h3; h4; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were not provided.A review of the device history records identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to the patient being non compliant to post op care as reported by the surgeon.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCR 6.5X50 SELF-TAP
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13415840
Report Number0001822565-2022-00264
Device Sequence Number14424750
Product Code LPH
UDI-Device Identifier00889024119857
UDI-Public(01)00889024119857(17)301204(10)J6909648
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K934765
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00625006550
Device Lot NumberJ6909648
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer03/24/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/05/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-