• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. SPEEDLOCK; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. SPEEDLOCK; BIT, DRILL Back to Search Results
Model Number 23-2010
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a hip labral repair surgery, the speed lock hip broke at the stopper.It snapped from the torque of the drill and hard bone.The procedure was completed with non-significant delay using the same device.Surgeon drilled what he could with the drill through the guide to start the hole, then free handed the rest since it was able to still mount on the chuck.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.The provided product identification information did not match any known release of this part and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDLOCK
Common Device Name
BIT, DRILL
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13415946
Report Number3006524618-2022-00040
Device Sequence Number4829207
Product Code HTW
UDI-Device Identifier00817470002935
UDI-Public00817470002935
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number23-2010
Device Catalogue Number23-2010
Device Lot Number20181031
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2022
Supplement Date Received by Manufacturer12/02/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-