• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC HAMMER GUIDE FOR TI ELASTIC NAILS; HAMMER, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC HAMMER GUIDE FOR TI ELASTIC NAILS; HAMMER, SURGICAL Back to Search Results
Catalog Number 359.218
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If the information is unknown, not available or does not apply, the section/field of the form is left blank.A product investigation was conducted.Visual inspection: the complaint device hammer guide f/ten (product code: (b)(4), lot number: unk) was returned to cq west chester for investigation.The hammer guide distal end had broken and the broken piece was not returned for investigation.The device also had impaction marks on its head and was severely deformed.Document /specification review: the date of manufacture for the item was not available, hence the current revision of drawing was reviewed.Dimensional inspection: complaint relevant dimension cannot be measured due to post manufacturing damage to the distal end.Conclusion: the hammer guide was broken and had a severe damage consistent with field usage.A definitive root cause was not determined during investigation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Lot number of the item was not available for review.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from depuy synthes reports an event in (b)(6) as follows: it was reported that during a procedure on (b)(6) 2021, the thread of the hammer guide was stripped.No further information provided.During manufacturer's investigation of the returned device it was also identified that the hammer guide distal end had broken and the broken piece was not returned for investigation.This device condition was evaluated and determined to be reportable on (b)(6) 2022.This report is for one (1) hammer guide for ti elastic nails this is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
If the information is unknown, not available or does not apply, the section/field of the form is left blank.A product investigation was conducted.Visual inspection: the complaint device hammer guide f/ten (product code: (b)(4), lot number: unk) was returned to cq west chester for investigation.The hammer guide distal end had broken and the broken piece was not returned for investigation.The device also had impaction marks on its head and was severely deformed.Document /specification review: the date of manufacture for the item was not available, hence the current revision of drawing was reviewed.Dimensional inspection: complaint relevant dimension cannot be measured due to post manufacturing damage to the distal end.Conclusion: the hammer guide was broken and had a severe damage consistent with field usage.A definitive root cause was not determined during investigation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Lot number of the item was not available for review.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from depuy synthes reports an event in (b)(6) as follows: it was reported that during a procedure on (b)(6) 2021, the thread of the hammer guide was stripped.No further information provided.During manufacturer's investigation of the returned device it was also identified that the hammer guide distal end had broken and the broken piece was not returned for investigation.This device condition was evaluated and determined to be reportable on (b)(6) 2022.This report is for one (1) hammer guide for ti elastic nails this is report 1 of 1 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAMMER GUIDE FOR TI ELASTIC NAILS
Common Device Name
HAMMER, SURGICAL
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK BALSTHAL (CH)
dornacherstrasse 20
balsthal 4710
SZ   4710
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key13416362
Report Number2939274-2022-00309
Device Sequence Number4870758
Product Code FZY
UDI-Device Identifier07611819104365
UDI-Public(01)07611819104365
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number359.218
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/28/2022
Initial Report FDA Received Date02/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-