(b)(4).The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.It was reported that the stent delivery system (sds) was inserted into the anatomy, removed, and then re-inserted.It should be noted that the xience skypoint instructions for use states (ifu); an unexpanded stent may be retracted into the guiding catheter one time only.An unexpanded stent should not be reintroduced into the artery once it has been pulled back into the guiding catheter.Subsequent movement in and out through the distal end of the guiding catheter should not be performed, as the stent may be damaged or dislodged during retraction back into the guiding catheter.The investigation determined the reported difficulties are related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|