| Model Number |
1012275-15 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a mildly calcified mid right coronary artery that is 75% stenosed.Once a non-abbott guide wire was advanced a 3.25x15mm trek balloon dilatation catheter (bdc) was attempted to be used for pre-dilatation; however, the bdc ruptured at 12 atmospheres during the second inflation.Another non-abbott device was used to pre-dilate the lesion, followed by stent deployment to complete the procedure.There was no adverse patient effects and no clinically significant delay was reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation determined the reported balloon rupture appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|