| Model Number |
20-000415 |
| Medical Device Problem Codes |
No Display/Image (1183); Misfocusing (1401); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial report ref to natus complaint# (b)(4).27 jan 2022 notification from (b)(6) of incident that occurred in (b)(6).Extension of the examination time, resulted in less effective sedation.It also created a work overload for the caregiver because of the need to process the images one by one to retransmit them for diagnosis by the doctor.Poor image quality for diagnosis.Impact on patient care.A questionnaire has been sent out to the customer.Further investigation to be carried out.The affected product to be evaluated.
|
| |
|
Event or Problem Description
|
|
Rohs compliant retcam shuttle.The customer reports that the device failed in the middle of an examination of a (b)(6) child, requiring the customer to get another device.The images were not transmitted to the pediatricians, forcing the caregiver to photograph the images with his phone, which resulted in loss of image quality.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up report 001 ref to natus complaint#: (b)(4).Requested return of the affected product.The system is expected to be shipped back on monday, 07 march 22.The returned affected product to be evaluated by natus.A questionnaire was sent out to the customer and it was confirmed that the customer does not want to answer the adverse event questionnaire.Per document (b)(4) rev r, risk assessment retcam, ref.Line - 9.1.5; hazard category/harm - delayed procedure; hazard - no imaging capability; hazard cause - component failure (laptop, computer, battery, power supply, etc); safety risk acceptable.
|
| |
|
Event or Problem Description
|
|
Rohs compliant retcam shuttle - the customer reports that the device failed in the middle of an examination of a 13-week old child, requiring the customer to get another device.The images were not transmitted to the pediatricians, forcing the caregiver to photograph the images with his phone, which resulted in loss of image quality.
|
| |
|
Additional Manufacturer Narrative
|
|
20 may 2022 - follow up report 001 ref to natus complaint# (b)(4).Requested return of the affected product.The system is expected to be shipped back next month.The returned affected product to be evaluated by natus.
|
| |
|
Event or Problem Description
|
|
Rohs compliant retcam shuttle - the customer reports that the device failed in the middle of an examination of a 13-week old child, requiring the customer to get another device.The images were not transmitted to the pediatricians, forcing the caregiver to photograph the images with his phone, which resulted in loss of image quality.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up report 003 ref to natus (b)(4).28-sep-2022 system was received in middleton.Engineering conducted the investigation.The failure was able to be duplicated.After about 10 minutes of viewing a live video image on the system, an error occurred that was caused by a loss of communication between the handpiece and the laptop.Communication to/from the handpiece was lost and the system froze.Reboot was necessary.After changing the usb port that the handpiece was connected to, the system functionality was restored and it continued to function correctly for over 24 hours.This intermittent usb communication issue is a known issue with some of the hp 650 model laptops.The handpiece was originally designed to work with usb 2.0.The hp 650 has only usb 3.0 ports.Most of the time the system works for long periods of time with no issues.A small percentage of these laptops seem to be more prone to this issue.There are no capa's related to this issue.This complaint does not identify a deficiency in the product design and therefore a capa is not required per risk assessment_18-000022, risk assessment retcam, ref.Line - 9.1.5, severity - serious, the risk is considered acceptable.A risk review is not required as this complaint does not describe a new failure mode or new harm and the existing hazard severity and/or probability of occurrence has not changed.Per qms-004442, complaint histories are reviewed routinely per quality system requirements and any complaint trends are assessed and documented as part of these reviews a device history record review is not applicable because, prior to the reported issue the device was in service for 2 or more years and a manufacturing defect is very unlikely.The system was sent in for repair in 2021 (b)(4).Correction to section d4 - udi number: (b)(4).Faillure confirmed:yes investigation result code :pleasanton|handpiece connection.Closure rationale: complaint verified, being tracked as a trend.Low safety risk of harm, individual complaint related to issue stated.Complaint will be included in trending data for further review and investigation if needed.
|
| |
|
Event or Problem Description
|
|
Rohs compliant retcam shuttle - the customer reports that the device failed in the middle of an examination of a 13-week old child, requiring the customer to get another device.The images were not transmitted to the pediatricians, forcing the caregiver to photograph the images with his phone, which resulted in loss of image quality.
|
| |
|