SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71335554
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994)
Date of Event 09/20/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
Us legal bilateral patient.It was reported that after a right hip revision surgery was performed on (b)(6) 2017 due to an infection, the patient kept suffering from an ongoing infection.One year after the revision, the patient was still taking suppressive antibiotics, as per indication of the infectious disease doctor.A nm bone scan 3 phase was performed on (b)(6) 2018 and suggested the presence of soft tissue inflammation.On (b)(6) 2019 the patient had an aspiration under general anesthesia and approximately 7 cc of somewhat cloudy joint fluid was removed from the right hip.It is unknown if/how the issue was treated.The outcome of the patient is unknown.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated infection.The clinical/medical investigation concluded that, the cloudy joint fluid may be consistent with the reported ¿chronic infection.¿ however, pathology was noted as negative for infection.With the information provided the clinical root cause of the cloudy joint fluid cannot be confirmed and it cannot be concluded that the reported clinical reaction was associated with a mal performance of the implant.No further documentation regarding any treatment related to the right hip was noted.It should be noted the patient was also being treated for a left hip infection.The patient impact beyond the aspiration cannot be determined.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for total hip systems revealed that postoperative deep wound infection and late periprosthetic infection has been identified as a potential complication.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the sterilization records revealed the batch was sterilized within normal parameters.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 54MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13416727
Report Number1020279-2022-00413
Device Sequence Number14625393
Product Code MBL
UDI-Device Identifier03596010598257
UDI-Public03596010598257
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K070756
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/15/2026
Device Model Number71335554
Device Catalogue Number71335554
Device Lot Number16DW08902
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer04/05/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date04/06/2022
Date Device Manufactured04/17/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SMF STEM WITH STIKTITE SZ 2-12MM06165
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age40 YR
Patient SexMale
-
-