| Model Number |
BEA28-110/I20-30 |
| Medical Device Problem Code |
Off-Label Use (1494)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Failure of Implant (1924)
|
| Date of Event |
01/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
|
| |
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa) with afx2 bifurcated stent graft, an afx vela infrarenal, an afx vela suprarenal, and an ovation ix extender.This case is outside the indications of use (off -label) due the use of afx with ovation devices.Approximately four (4) years post initial procedure a type ib endoleak was identified.The patient was asymptomatic.Reintervention was completed with embolization of the hypogastric and extension of the left iliac down to the external iliac.First an ovation ix iliac limb (14x100) was deployed and then the afx limb was extended with deployment of an ovation ix extender (22x45).The type ib endoleak was successfully resolved.The patient was reported to be fine post-secondary procedure.
|
| |
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa) with afx2 bifurcated stent graft, an afx vela infrarenal, an afx vela suprarenal, and an ovation ix extender.This case is outside the indications of use (off -label) due the use of afx with ovation devices.Approximately four (4) years post initial procedure a type ib endoleak was identified.The patient was asymptomatic.Reintervention was completed with embolization of the hypogastric and extension of the left iliac down to the external iliac.First an ovation ix iliac limb (14x100) was deployed and then the afx limb was extended with deployment of an ovation ix extender (22x45).The type ib endoleak was successfully resolved.The patient was reported to be fine post-secondary procedure.Additional information received per clinical assessment indicating that "moderate" aaa sac growth also occurred at time of reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the reported type ib endoleak (of the left common iliac artery) and secondary endovascular procedure are confirmed.This is consistent with the reported adverse event/incident.The clinical evaluation also shows reasonable evidence to suggest aneurysm sac enlargement (unknown diameter) occurred that was not included in the reported event.This finding was discovered during an examination of the operative note dated 21 january 2022.Procedure-related harms, device, user, procedure, or anatomy relatedness of this event could not be determined with the medical records available for review.The final patient status was reported as being discharged on the first post operative day at home in good condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections: h6 investigation finding codes; remove 3233.H6 investigation conclusion codes; remove 11.
|
| |
|