WRIGHTS LANE SYNTHES USA PRODUCTS LLC SPEED TITAN 15X15X15MM IMPLANT KIT; STAPLE, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
SE-1515TI |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, the circulating nurse opened a 15x15 speed, titan staple, to the back table.The staple was not connected to the inserter when opened.The doctor and pa tried to reload the staple but could not.A new 15x15 staple was opened.There was a five (5) minute surgical delay.The procedure successfully completed.The patient outcome is unknown.This report involves one (1) speed titan 15x15x15mm implant kit.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Expiration date part # se-1515ti synthes lot # bse170236 supplier lot # n/a release to warehouse date: 25 apr2017 expiration date: 20 mar2022 manufactured by: bme no ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the speedtitan compression implant kit 15x15 has a broken section from one of the tips of the inserter, assembling issues are most likely due to this condition; fragment and staple are included in the package.Additionally, a cracking condition is visible also on the tip.A capa was initiated for this type of nonconformance, the most important ones being the design of the stick and the molding process at the supplier.A dimensional inspection for the speedtitan compression implant kit 15x15 was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces as mentioned in the event description.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the speedtitan compression implant kit 15x15 would contribute to the complained device issue.This is a known issue with se-1515ti.After investigation, it was determined that there were several contributing factors for this type of nonconformance, the most important ones being the design of the stick and the molding process at the supplier.Although the injection molding process was identified as the most likely root cause, the supplier stated they could not adhere to bme requirements, therefore capawas initiated to document the migration from supplier protolabs to supplier apex.Corrective actions (design changes and mold update) have been implemented.No manufacturing related issue was identified and/or confirmed, therefore review to the specific prm and prm line is not required.The potential impact of the design in the complaint condition has been addressed under a capa.No further corrective actions are required at this moment.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|