MAUDE Adverse Event Report: DEPUY MITEK LLC US CLEARCANNULA-THREADED 8.5MMX75MM 5PK -ST; ARTHROSCOPIC DRAINAGE CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US CLEARCANNULA-THREADED 8.5MMX75MM 5PK -ST; ARTHROSCOPIC DRAINAGE CANNULA Back to Search Results
Model Number 214120
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer in (b)(6) that during a shoulder repair surgery on (b)(6) 2022, it was observed that the rubber seal was defective on the clear cannula-threaded 8.5mmx75mm device that it came loose and floated around in the situs.It was reported that through extensive searching it was recovered.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
Additional narrative: udi: (b)(4).Incomplete.The expiration date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H4: the device manufacture date was reported as unknown on the initial report; and has been updated accordingly.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation, however the customer provided a photo, upon visual inspection of the photo, it could be observed that the cannula is showing the silicone seal detached.A manufacturing record evaluation was performed for the finished device 2103222 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the inspection of the photo, this complaint can be confirmed.The possible root cause for reported failure can be attributed to procedural variables, such handling of the device or product interaction during procedure; excessive force being applied to pass instruments through the cannula which would cause a detachment of the silicone seal, however, this cannot be conclusively determined.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARCANNULA-THREADED 8.5MMX75MM 5PK -ST
Common Device Name
ARTHROSCOPIC DRAINAGE CANNULA
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key13417753
Report Number1221934-2022-00364
Device Sequence Number3872043
Product Code GEA
UDI-Device Identifier10886705003617
UDI-Public10886705003617
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Model Number214120
Device Catalogue Number214120
Device Lot Number2103222
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2022
Supplement Date Received by Manufacturer02/27/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date02/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-