| Model Number |
3921848 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the connector is small and the catheter can not enter.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample was returned for evaluation.The complainant's returned unit was visually inspected.The port was assembled with an incorrect component which was smaller in size.Potential root causes were: during kitting of clearance sleeves, mislabeling occurred, and the operator did not detect the discrepancy or during returns both part number of the sleeves were mixed.A quality alert was issued.Awareness notification was disseminated to personnel in order to explain the importance to adherence or following in the procedure.A new visual aid was created for personnel to inspect sleeves before assembly on 21-sep-2022.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
|
| |
|