• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION VECTRA VASCULAR ACCESS REVISION GRAFT; PROSTHESIS, VASCULAR GRAFT, OF LESS THEN 6MM DIAMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION VECTRA VASCULAR ACCESS REVISION GRAFT; PROSTHESIS, VASCULAR GRAFT, OF LESS THEN 6MM DIAMETER Back to Search Results
Model Number 10002-5020-003
Medical Device Problem Codes Crack (1135); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/28/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that after the tunneler was removed, the graft was cracked during the pulling to adjust the position in the blood vessel.
 
Additional Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Additional Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Event or Problem Description
It was reported that after the tunneler was removed, the graft was cracked during the pulling to adjust the position in the blood vessel.
 
Additional Manufacturer Narrative
Manufacturer's investigation conclusion: the report of damage to the vectra vascular access graft was confirmed during the investigation of the returned graft segments.Two segments of graft material were returned.The longer segment, measuring approximately 3.5in, appeared unremarkable.The shorter segment showed multiple splits in the outer layer of the graft.The innermost layer of the graft was visible through the splits.The cut end of the shorter segment was also noted to be unraveling.The undamaged areas of the graft segments were gently twisted, bent, and stretched, which did not generate additional splits in the outer layer.Although the specific root cause of the splits in the outer layer could not be conclusively determined, it was reported that the graft was pulled to adjust positioning, after the tunneler was removed.The vectra vascular access graft ifu states that the graft should not be pulled (axially elongated) or stretched during handling at implantation.Excessive elongation or stretching of the graft will damage the microporous layers of the graft.Review of the device history records showed no deviations from manufacturing or qa specifications, including visual and dimensional inspection.Graft preparation and implantation are addressed in the vectra vascular access graft (vag) instructions for use (ifu).The ifu states that the graft should not be pulled (axially elongated) or stretched during handling at implantation.The graft should be trimmed long enough to prevent stress on the anastomosis and allow for a full range of body motion when implanted.Excessive elongation or stretching of the graft will damage the microporous layers of the graft.The ifu states that the vectra sheath tunneler should be used to minimize subcutaneous trauma and the force required to position the graft during implant.Tunneling permits graft placement without pulling - which can produce excessive forces that result in damage to the graft microporous layers.When positioning and trimming the graft, avoid those areas of less dense reinforcement, especially near the anastomotic end.The non-reinforced segment must be trimmed to allow for proper sizing of the graft.The tunneling section of the ifu states that the vectra vag tunnelers (except for the full curve tunneler which can only be used the 5mm) can be used with either the 5mm or 6mm vectra vag.Liberally irrigate the inside of the tunneler sheath as well as all surfaces of the graft with a sterile solution of normal physiologic saline to facilitate slipping the graft through the tunneler.Care should be taken to push rather than pull the graft through the tunneler.The vectra vag can be implanted using either a one-sheath or two-sheath technique and instructions for each are provided.These instructions includes steps that that ¿insert the vectra vag into the subcutaneously placed sheath and gently push the graft through using a rotating motion if necessary, while irrigating liberally with sterile saline.¿ no further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VECTRA VASCULAR ACCESS REVISION GRAFT
Common Device Name
PROSTHESIS, VASCULAR GRAFT, OF LESS THEN 6MM DIAMETER
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key13418269
Report Number2916596-2022-00390
Device Sequence Number3963255
Product Code DYF
UDI-Device Identifier00813024011316
UDI-Public00813024011316
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K001927
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/30/2024
Device Model Number10002-5020-003
Device Catalogue Number10002-5020-003
Device Lot Number7173640
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer04/05/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date04/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/30/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-