Model Number MMT-105NNBLNA |
Device Problem
Computer Software Problem (1112)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 01/04/2022 |
Event Type
malfunction
|
Event Description
|
Information received by medtronic indicated that there was an inaccuracy between the dose intended and dose recorded.The dose was greater than 1.0 unit of insulin.Customer¿s blood glucose was unknown.No harm requiring medical intervention was reported.The insulin pen will be returned for analysis.
|
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Manufacturer Narrative
|
Serial number: n/a software version: n/a color: blue battery life remaining: n/a customer reports: inpen app is not recording the exact doses dialed on the inpen.Per visual inspection: no physical damage was noted to inpen.The inpen does not pair with commercial mobile app.Inpen received with leadscrew 1/4 of the travel.Re-wound screw.No drag was observed, the screw was not bent, advanced when dosage knob was pressed dialing a dosage and retracted appropriately.No resistance was observed when dosing without a cartridge installed.The screw advanced every time 30.0u was dialed and dosed until the screw reached max extension.In conclusion: the customer complaint of inpen app not recording exact doses dial on the inpen could not be confirmed due to inpen not pairing to inpen app.Clinical - destructive testing showed that leadscrew anomaly was caused by a pattern wheel misalignment due to an encoder base bond failure.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Color: blue customer reports: inpen app is not recording the exact doses dialed on the inpen.Per visual inspection: no physical damage was noted to inpen.The inpen does not pair with commercial mobile app.Inpen received with leadscrew 1/4 of the travel.Re-wound screw.No drag was observed, the screw was not bent, advanced when dosage knob was pressed dialing a dosage and retracted appropriately.No resistance was observed when dosing without a cartridge installed.The screw advanced every time 30.0u was dialed and dosed until the screw reached max extension.In conclusion: the customer complaint of inpen app not recording exact doses dial on the inpen could not be confirmed due to inpen not pairing to inpen app.Per (b)(6), clinical - destructive testing showed that leadscrew anomaly was caused by a pattern wheel misalignment due to an encoder base bond failure.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|