Model Number 100/111/025
Medical Device Problem Code Device Damaged Prior to Use (2284)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was broken plastic upon arrival.
 
Additional Manufacturer Narrative
End user: (b)(6).
 
Additional Manufacturer Narrative
Other text: (b)(4) samples were returned for evaluation; they were in used condition.The returned samples were visually inspected; broken connectors were observed in both samples.The complaint was confirmed.The root cause for the fractured components is undetermined; it is believed that the damage occurred after leaving the facility.Customer complaint notification was performed on 27-jun-2022 to personnel to explain the reported failure mode.No lot number was provided for performance of a device history record review.No patient injury was reported.(updated b5 and h6), corrected data: labeled for single use was answered incorrectly, in the initial report.
 
Event or Problem Description
There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Common Device Name
TUBE TRACHEAL
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
rm401, (gems bldg, gunja-dong)
minneapolis, MN 55442
MDR Report Key13419032
Report Number3012307300-2022-02313
Device Sequence Number4826577
Product Code BTR
UDI-Device Identifier15019315018848
UDI-Public15019315018848
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K790312
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100/111/025
Device Catalogue Number100/111/025
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date10/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-