Details of complaint: the biomedical engineer (bme) reported that the central nurse's station (cns) had shut itself down when the monitor technician clicked the "discharge" button to move a patient from a wired bedside monitor (bsm) to a transmitter.No patient harm was reported.Investigation summary: as no devices were returned for evaluation relating to this event, the reported issue could not be duplicated nor confirmed.As such, a root cause cannot be determined.The cns shutting down is commonly a symptom of failing internal components, commonly the hdds.The device was installed in 2011.The hdds are electronic components that may deteriorate through time and may wear from continual use.Possible causes of failure include mechanical failure from improper use/handling, corruption of the files from abnormal shutdowns, viruses, or power issues.Other possible causes of the issue are power surges/interruptions and wear and tear.Any events that involve fluctuations and changes in the voltage such as power outages and generator tests may damage the hdds.Fluid intrusion, dust accumulation, or a lack of maintenance on fans, air intakes, and other components may lead to overheating of the hard drive and subsequent failure.When hard drives fail, this failure can manifest in different ways.There can be an error message (e.G., "hdd access error," "ssd access error," "ssdx error," "threshold breach"), or the device may reboot, display a blue failure screen, or the device screen could "freeze." the unit may sometime then restart, or the device may need to have the hard drive replaced.Sometimes the unit can be rebuilt by the backup disk (i.E., the system has redundant array of independent disks (raid), which puts multiple hard drives together to improve performance).Nk will continue to monitor and trend similar complaints.Due to the age of the device and lack of servicing history, the cns was likely failing due to hdd failure as a result of wear and tear.The following fields contain no information (ni), as attempts to obtain information were made, but not provided: a2 - a6 b6 b7 d10 attempt # 1: (b)(6)2022 emailed the customer via microsoft outlook for all information in the ni list: no reply was received.Attempt # 2: 01/18/2022 emailed the customer via microsoft outlook for all information in the ni list: no reply was received.Attempt # 3: (b)(6)2022 emailed the customer via microsoft outlook for all information in the ni list: no reply was received.Additional device information: d10 concomitant medical device: the following devices were used in conjunction with the cns; however the model and serial numbers are listed as no information (ni) as attempts to obtain the information were made but not provided.Bedside monitor (bsm): model #: ni serial #: ni transmitter: model #: ni serial #: ni additional information: b4 date of this report g3 date received by manufacturer g6 type of report h2 if follow-up, what type? h6 event problem and evaluation codes h10 additional manufacturer narrative.
|