| Model Number |
4720 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchors remained loose in the joint.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for investigation and the reported failure mode was not confirmed.Alleged failure: thread broken.Probable root cause: design: inadequate raw material specification (needle or depth indicator); needle tube thickness too thin or too thick; suture does not have adequate working length; inadequate handle assembly design ; inadequate raw material ; depth indicator thickness too thin or too thick.Process: depth indicator not manufactured to specification; needle or implant body not manufactured to specification; dimple process on needle not manufactured to specification; inadequate design of needle/depth indicator to create interference to retain/remove depth indicator; stiffener not manufactured to specification.Application: user intentionally bends/forces instrument (g8); incision too small (g9); inadequate or improper visualization (g10); user not familiar with device use (g11); user does not use sled or other technique to prevent catching on soft tissue (g12).The device manufacture date is not known.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchors remained loose in the joint.
|
| |
|