MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE AIR MENISCUS SYSTEM, CURVED; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE AIR MENISCUS SYSTEM, CURVED; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number 4720
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the anchors remained loose in the joint.
 
Additional Manufacturer Narrative
The product was not returned for investigation and the reported failure mode was not confirmed.Alleged failure: thread broken.Probable root cause: design: inadequate raw material specification (needle or depth indicator); needle tube thickness too thin or too thick; suture does not have adequate working length; inadequate handle assembly design ; inadequate raw material ; depth indicator thickness too thin or too thick.Process: depth indicator not manufactured to specification; needle or implant body not manufactured to specification; dimple process on needle not manufactured to specification; inadequate design of needle/depth indicator to create interference to retain/remove depth indicator; stiffener not manufactured to specification.Application: user intentionally bends/forces instrument (g8); incision too small (g9); inadequate or improper visualization (g10); user not familiar with device use (g11); user does not use sled or other technique to prevent catching on soft tissue (g12).The device manufacture date is not known.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
 
Event or Problem Description
It was reported that the anchors remained loose in the joint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIR MENISCUS SYSTEM, CURVED
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key13419498
Report Number0002936485-2022-00061
Device Sequence Number4867213
Product Code GAT
UDI-Device Identifier07613327375404
UDI-Public07613327375404
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K153087
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4720
Device Catalogue Number4720
Device Lot Number7002913
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/06/2022
Supplement Date Received by Manufacturer01/06/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-