STRYKER GMBH CONNECTING NUT - SHORT HOFFMANN LRF M8 X 6MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 4933-1-010
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
The following event was reported by the stryker rep: "the hoffmann lrf frame was put on the patient on (b)(6) 2021 for an ankle fusion procedure to protect the internal hardware.Patient notified the surgeon on (b)(6) 2022 complaining of a loud 'pop' or a loose 'clicking sound.Frame was removed on the following day (b)(6) 2022.Every wire broke and nuts loosened up.The surgeon mentioned the patient was not walking on it throughout the few weeks it was on.".
 
Additional Manufacturer Narrative
Please note corrections to section d9/h3, the device was lost and could not be returned.The reported event could be confirmed, since the product was visible missing on the returned hoffmann lrf construct.The inspection revealed the following: on the foot ring (left side, in the front), we were able to locate one missing nut from one of the four telescopic strut.The threaded bolt shows no deformation or damage.Furthermore, it is visible at the foot ring, that a nut was attached, as the foot ring is slightly discolored at this spot were the nut was.Based on this, it can only by that the nut didn't get tightened enough.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a use related issue.If the device is returned or if any additional information is provided, the investigation will be reassessed.H3 other text : device disposition is unknown.
 
Event or Problem Description
The following event was reported by the stryker rep: "the hoffmann lrf frame was put on the patient on (b)(6)21 for an ankle fusion procedure to protect the internal hardware.Patient notified the surgeon on (b)(6)-22 complaining of a loud 'pop' or a loose 'clicking sound.Frame was removed on the following day (b)(6)-22.Every wire broke and nuts loosened up.The surgeon mentioned the patient was not walking on it throughout the few weeks it was on.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTING NUT - SHORT HOFFMANN LRF M8 X 6MM
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13419620
Report Number0008031020-2022-00029
Device Sequence Number17312431
Product Code KTT
UDI-Device Identifier07613252610304
UDI-Public07613252610304
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K113327
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number4933-1-010
Device Catalogue Number49331010
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/06/2022
Supplement Date Received by Manufacturer03/25/2022
Initial Report FDA Received Date02/01/2022
Supplement Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexMale
-
-