• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT MEDICAL SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT MEDICAL SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 8CN85ED
Medical Device Problem Codes Defective Component (2292); Gas/Air Leak (2946)
Health Effect - Clinical Code Unintended Extubation (4564)
Date of Event 12/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Uf/importer rpt num: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the respiratory therapist was paged to the patient and found the cuff on the patient's tube was deflated and had an audible cuff leak.Attempts to inflate the balloon were unsuccessful, would quickly deflate.The original placing provider team notified and exchanged for a substitute trach at the bedside.The md chose a different tube brand because that was the second trach change in 24 hours due to a cuff leak.When the tube was removed and submerged in water, a leak through a hole in the cuff was noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
EI  3810
Manufacturer (Section G)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
EI   3810
Manufacturer Contact
clay chandler
5920 longbow drive
boulder, CO 80301
3035306409
MDR Report Key13420095
Report Number8020889-2022-00025
Device Sequence Number4854039
Product Code JOH
UDI-Device Identifier10884521726239
UDI-Public10884521726239
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K193077
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2026
Device Model Number8CN85ED
Device Catalogue Number8CN85ED
Device Lot Number202102080X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2022
Initial Report FDA Received Date02/01/2022
Date Device Manufactured02/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
Patient Weight123 KG
-
-