• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number DIU150
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the intraocular lens (iol) was fully inserted and removed during the same procedure when after the insertion the surgeon discovered the iol was cracked.The replacement lens is the same model and diopter.There was no patient injury reported.No further information is available.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the product was not returned for evaluation.The reported complaint cannot be confirmed.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.An attempt has been made to obtain missing information; however, to date, the information has not been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Section d10: device available for evaluation: yes.Section d10: returned to manufacturer on: 2/8/2022.Device evaluation: the lens was received cut in half and lens damage was identified on the optic body.No defects or assembly issues were observed with the handpiece.The complaint issue was confirmed, however this can be attributed to excess ovd (residue observed inside the lens module) usage during loading and insertion which can compromise lens placement in the staging area and contribute to deployment issues, and therefore no product deficiency could be identified.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13421275
Report Number3012236936-2022-00277
Device Sequence Number4854075
Product Code MJP
UDI-Device Identifier05050474746541
UDI-Public(01)05050474746541(17)241109
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2024
Device Model NumberDIU150
Device Catalogue NumberDIU150U200
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2022
Supplement Date Received by Manufacturer03/11/2022
Initial Report FDA Received Date02/02/2022
Supplement Report FDA Received Date03/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-