MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH-AT ARTIS ZEE CEILING; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH-AT ARTIS ZEE CEILING; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10094137
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the artis zee ceiling system.During an interventional procedure, the user reported that the cable of the table controller (tcm) attached to the sleeper became caught in the ceiling stand.As a result, the cable was broken and the stand could no longer be operated.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the artis zee ceiling system.During an interventional procedure, the user reported that the cable of the table controller (tcm) attached to the sleeper became caught in the ceiling stand.As a result, the cable was broken and the stand could no longer be operated.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be user error.The investigation was performed considering complaint description, cs reports, system history, and system log files.During a procedure, the cable to the control module was caught in the cover of the ceiling stand and during subsequent movement of the c-arm the cable was ruptured.The emergency stop circuit runs through the cable as well resulting in an interruption of the emergency stop circuit.This interruption of the emergency circuit lead to a block of all movements.X-ray was still available and was not affected by this issue.The customer service engineer (cse) investigated the system and found the cable was defective.Further information was provided that the user was aware that the cable was caught in the cover of the ceiling stand but decided to move the system anyway resulting in the ruptured cable.The cable was replaced and the system works as intended afterwards.Customer inattention could be determined as the relevant root cause.There were no further issues reported and the system works as intended.A possible general fault that would require corrective measures of the installed base could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE CEILING
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
4843231631
MDR Report Key13421493
Report Number3004977335-2022-14382
Device Sequence Number6009821
Product Code OWB
UDI-Device Identifier04056869010052
UDI-Public04056869010052
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181407
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Remedial Action Repair,Inspection
Type of Report Initial,Followup
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10094137
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/26/2022
Supplement Date Received by Manufacturer04/07/2022
Initial Report FDA Received Date02/02/2022
Supplement Report FDA Received Date04/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-