• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD VERSAPORT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD VERSAPORT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number ONB12SHF
Medical Device Problem Codes Leak/Splash (1354); Decrease in Pressure (1490)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic gastrectomy, pneumoperitoneum leak occurred and the pneumoperitoneum could not be maintained on the five devices.It was noted air or gas leakage from the seal occurred at the first 12 mm trocar, then it was also occurred to other trocars.At the end of the surgery, a pneumoperitoneum leak sound was heard from the 5 mm trocar.There was a rapid loss of pneumoperitoneum noted.A new product was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.Three devices were available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that there was rapid loss of pneumoperitoneum caused by device failure and the device was leaking fluid or air.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPORT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key13422709
Report Number9612501-2022-00154
Device Sequence Number20280144
Product Code GCJ
UDI-Device Identifier10884521181366
UDI-Public10884521181366
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K130435
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberONB12SHF
Device Catalogue NumberONB12SHF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2022
Supplement Date Received by Manufacturer03/14/2022
Initial Report FDA Received Date02/02/2022
Supplement Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-