As reported, during an angiogram of the popliteal artery the hub of a advance micro 14 ultra low-profile pta balloon catheter broke.The balloon was advanced through another manufacturer's 6fr sheath to the target location.It was inflated using an advantage kit with a 50/50 solution of saline and contrast.The hub of the device then broke while "twisting".The balloon was then removed as a unit and the procedure was completed using another balloon.It was noted that the anatomy was not tortuous, but was moderately to severely calcified.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.The patient did not experience any adverse effects due to this occurrence.
|
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, during an angiogram, the hub of a advance micro 14 ultra low-profile pta balloon catheter broke.The balloon was advanced through another manufacturer's 6fr sheath to the target location.It was inflated using an advantage kit with a 50/50 solution of saline and contrast.The hub of the device then broke while "twisting".The balloon was then removed as a unit and the procedure was completed using another balloon.It was noted that the anatomy was not tortuous, but was moderately to severely calcified.The severely occluded lesion was located in the superficial femoral/popliteal artery.Mild stenosis was noted in the profunda femoris artery, and severe stenosis was reported in the adductor hiatus and infrapopliteal area.An unspecified 0.014-inch wire was in the lumen of the device at the time of the event.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record, drawings, instructions for use (ifu), and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complainant returned one advance micro ¿ 14 ultra low-profile pta balloon catheter to cook for investigation.Physical examination of the returned device revealed the hub was cracked and broken, but not separated.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with the complaint device lot.Because there were no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in the field.The instructions for use (ifu) provides the following information to the user related to the reported failure mode: "upon removal from package, inspect the product to ensure no damage has occurred." cook reviewed the device master record (dmr) and product labeling, including the ifu, to confirm adequate manufacturing specifications, operation parameters, and the design verification testing of the device are in place to identify and prevent this failure mode prior to distribution.Cook concluded that a component failure unrelated to the design or manufacturing of the complaint device caused this incident.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|