• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97715
Device Problems Overheating of Device (1437); Inappropriate/Inadequate Shock/Stimulation (1574); Connection Problem (2900)
Patient Problems Burn(s) (1757); Electric Shock (2554)
Event Date 01/27/2022
Event Type  malfunction  
Event Description
Information was received from a patient (pt) who was implanted with an implantable neurostimulator (ins) for spinal pain.It was reported that their recharger was getting so hot it was burning them (pt was not sure if it had been leaving a mark).Pt mentioned that the cord was also bubbling.An email was sent to the repair department to replace the recharger.
 
Manufacturer Narrative
Concomitant medical products: product id: 97755, serial#: (b)(4), product type: recharger.Other relevant device(s) are: product id: 97755, serial/lot #: (b)(4), ubd: 28-oct-2019, udi#: (b)(4).Date of event: month and year valid.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Concomitant medical products: product id 97755, lot#/serial# (b)(6) product type recharger.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received.It was reported the circumstances that led to the recharger hot and burning was it appeared that there was a small crack in the wiring where the paddle neck and wire meet which caused a short, which in turn caused the wire to get hot and burn/shock the patient's skin.The issue was resolved, the patient was sent a new paddle which completely resolved the problem.
 
Manufacturer Narrative
H3: product id 97755, serial# (b)(6), was returned for product analysis.Analysis found the complaint was unverified, the recharger telemetry module (rtm) never got hot after running for 10 minutes.No anomalies were visually observed on the rtm.The device was scrapped due to a t6030 failure (rms voltage at the h probe), there was a suspected recharge coil that was defective.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
See h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13423450
MDR Text Key284872915
Report Number3004209178-2022-01513
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/28/2019
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/01/2022
Initial Date FDA Received02/02/2022
Supplement Dates Manufacturer Received02/25/2022
05/16/2022
Supplement Dates FDA Received03/08/2022
05/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Age43 YR
Patient SexFemale
Patient Weight84 KG
-
-